Air-Tight X-Ray Radiation Protection Door

Modes / Features Supported
Automatic mode √
Air leakage prevention mode √
Hermetic sliding mode √
Manual (hand) open/close √
Foot (kick) open/close √
Safety mode to prevent door pinch √
Speed control for open/close √
Door locking √
Emergency backup power in case of blackout √
X-ray protection (optional) √
Door load capacity up to 180 kg √

Automatic Air-Tight Lead Door

The door is controlled via modern technology. It uses a belt drive automatic system with a motor imported from Germany that meets international quality standards.

This hermetic (air-sealed) lead door is 50 mm thick and includes a full rubber seal around the perimeter of the panel. The lead cladding is 2 mm thick. If there is a viewing window, it is made of radiation-shielding glass.

Components of the Automatic Lead Door System

  1. Sliding rail track (Hermetic type) that slopes to seal the door panel against the frame
  2. 100 W drive motor capable of handling heavy X-ray doors (150–200 kg)
  3. Controller for setting door behavior (speed, open/close timing, safety features)
  4. Durable timing belt (imported from Italy) to drive the door
  5. High-quality hangers/rollers made of Nylon6, load rating up to 200 kg
  6. Idle pulley
  7. Sensors for detecting and controlling door movement
  8. Safety beam to prevent door pinch

Customer Options

Customers can choose from:

  • Sliding lead radiation protection door
  • Swing (hinged) lead radiation protection door
  • Automatic lead radiation protection door

Lead Thickness Options by Room / Use

Room / Equipment Type Suggested Lead Thickness for Door
Dental X-ray 1.0 mm Pb
Mammogram / Bone Density Low energy → lower thickness
General X-ray 1.5 mm Pb
CT Scan 2.0 – 3.0 mm Pb
Fluoroscopy 2.5 mm Pb
Cath Lab / Hybrid OR Higher shielding (up to ~3 mm)
ERCP / Intervention Rooms Higher shielding requirement

Examples of test reports are cited:

  • General X-ray: 1.5 mm (report 46201022021)
  • CT Scan: 2 mm (report 0463167380)
  • Fluoroscopy: 2.5 mm (report 0463346807)
  • Hybrid / Intervention / ERCP: up to 3 mm (report 04970003851)

Semi-Airtight Door (Option)

Specification
Double glazing (double glass) √
Maximum panel width: 120 cm √
Interior filled with fire-resistant polyurethane foam √
Equipped with socket bolts for drop seal adjustability √
Overall panel thickness ≥ 50 mm √
Synthetic rubber seals on three sides √
Bottom edge with automatic drop seal √
Lead cladding for the full panel √
Suitable for operating rooms, X-ray rooms, wards, etc. √
Location applicable
Operating rooms √
X-Ray rooms √
WARDS √