GRAMLOCK — X-Ray Shielding Board

GRAMLOCK X-ray shielding board is a rigid panel that can effectively block X-ray radiation without using lead (Lead Free). It is manufactured from gypsum (dihydrate gypsum, CaSO₄·2H₂O) and nonwoven glass fiber fabric. It has strength, fire insulation, and contains barium sulfate (BaSO₄) which provides X-ray shielding properties that are safe for living organisms and the environment (a green building material).

Technical Specifications

  • Thickness: 14 mm
  • Panel size: 1,220 × 2,440 mm
  • X-ray shielding performance equivalent to 1.0 mm lead (Pb)
  • Note: Tested & certified by the Department of Medical Sciences

Features of GRAMLOCK Lead-Free Board

  • Shields against X-ray radiation from X-ray generators and sources, comparable to lead (Pb)
  • Non-flammable (Test method NM-3472)
  • Acts as a fire insulation (thermal barrier)
  • Green / ecological X-ray shielding product, free from lead
  • Certified by the Department of Medical Sciences, Ministry of Public Health, test report No. 0460005683 for 120 kVp showing lead equivalent 1.0 mmPb
  • For lead equivalent of 2.0 mmPb, use double layering
  • Another test report No. 0463346808 shows lead equivalent 2.21 mmPb with 99.94% attenuation
  • Durable — resistant to water and moisture
  • Manufacturer holds ISO 9001:2008 / JIS Q 9001:2008 certification

Recommended X-ray Radiation Shielding (by Room Type)

Equipment / Room Type Recommended Shielding (Lead Equivalent) Note
Dental X-ray / Mammography / Bone Densitometry 1.0 mmPb KVP values must be verified for each installation.
General X-ray Room 2.0 mmPb (2 layers)
Hybrid Operating Room (OR) 2.0 mmPb (2 layers)
CT Scan Room 3.0 mmPb (3 layers)
Angiography Room 3.0 mmPb (3 layers)
Fluoroscopy Room
Cardiac Catheterization Laboratory (Cath Lab)

Installation Method

Gram Lock wall, Type 1 (single-sided), lead equivalent thickness: 1.0 mmPb

Gram Lock wall, Type 1 (single-sided), lead equivalent thickness: 2.0 mmPb